Chapter 6: Maintenance and Troubleshooting
FARO Laser Line Probe Compensation
Ensure the Scan Rate and Scan Density are set 1/1. In the Probe
Management dialog box, edit the current Laser Line Probe. to access the
FARO L
ASER LINE PROBE CONTROL dialog box.
Perform the compensation at least 3 times. See “FARO Laser Line
Probe Compensation” on page 97.
When sweeping from side to side and front to back, make sure you
sweep through at least 90 degrees.
Typically, you should take about 200 scans during the compensation
process.