Chapter 3: Operation
Diagnostic Tilt
Click DIAGNOSTIC TILT to check for any tilt in the base of the Quantum. This
command uses the Quantum’s internal inclinometer.
1. Move the tubes and joints of the Quantum.
• Any tilt of the base is recorded and displayed on the screen. All of the
points must be within the circle on the graph.
•Click Reset to clear the screen.
• Rotate the mouse wheel to zoom in and out of the chart. Right-click and
drag to move the chart around the window.
•Click RESET PLOT AXES or double-click the wheel button to reset the
chart view.
2. Click Close to exit.
Figure 3-50 Diagnostics dialog box