Chapter 5: FiberInspector™ Pro Tests
FI-3000 Probe Setup
109
With an adapter attached to the tip, to see the correct numbers
for the fibers in the test results:
• For an opposed key adapter, set the Key Position to the key
position on the Probe.
• For an aligned key adapter, set the Key Position to the
opposite side from the key position used on the Probe.
With Key Position set correctly for the type of adapter, fiber
number 1 shows on the top left of the image on the display.
Note
If you rotate the adapter or tip on the probe,
change the key position to Key Left or Key Right,
as appropriate to maintain fiber number
agreement.
Figure 45 shows the key on the left side of the Probe and the Key
Position set to Key Left on the tester to test connectors in a patch
panel with opposed key adapters.
www.GlobalTestSupply.com