Versiv Series Cabling Certification Product Family
Users Manual
42
This icon shows when the talk function is on. To use the
talk function:
1 Connect the main and remote testers together through a
link that has one or more good wire pairs.
2 Connect headsets to the headset jacks on the testers.
3 Press the button on one of the headset microphones or
press on the remote, then speak into the
microphone.
Make Sure Your Tester is Ready to Certify
Cabling
To make sure your tester meets its accuracy specifications, follow
these guidelines:
Keep the tester’s software current. The latest software is
available on the website. See “Update the Software” on
page 359.
Set the reference for the twisted pair adapters every 30 days.
See Set the Reference below.
Make sure that you select the correct cable type for the job, and
that the NVP for the cable is correct. See Table 3 on page 46.
Make sure you select the correct test limit for the job. See
Table 3 on page 46.
Make sure the cords and connectors for all test equipment
and patch cords are in good condition.
Make sure the battery is fully charged.
Send the modules to a service center every 12 months for
factory calibration.
www.GlobalTestSupply.com