Versiv Series Cabling Certification Product Family
Users Manual
70
The location of the measurements. To switch between results
for the main and remote, tap REMOTE or MAIN (
).
Measured values for the wire pairs.
The limit line (in red) for the measurement.
Note
If the limit line is black, the tester does not
evaluate the measurement at those frequencies
because a dB rule applies. See the Technical
Reference Handbook.
The vertical scale is the measured value in decibels.
The horizontal scale is the frequency range in megahertz.
To see help for the screen, tap .
To switch between results for the main unit and the remote,
tap REMOTE or MAIN.
The margin at the cursor’s location. The margin is the
difference between the measured value and the limit. The
margin is negative if the pair failed.
The measured value at the cursor’s location.
When you first look at the plot, the cursor is at the frequency
of the worst margin. To move the cursor to the worst value,
tap WORST VALUE. (The insertion loss plot is different. See
the Technical Reference Handbook.)
To see the plots for pairs, tap or . To select pairs to
show on the plot, touch or for one second to see the
SELECT PAIRS window. Select the pairs you want to see, then
tap OK.