EasyManua.ls Logo

Gantner Q Series - Voltage; Current

Gantner Q Series
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vers. No. 6.1
88 Released: 25/04/2017
4 Connecting the modules Q.bloxx A127: Connecting sensors
Tip
Further information on transducers and sensors can also be
found in Chapter 6 ff. page 135.
4.21.1 Voltage
You can measure voltage differences of up to ±1200 V. Here, var-
ious input voltage ranges from ±40 V
DC
to ±1200 V
DC
are possi-
ble. The voltage level (the potential) may be up to 1200 V
DC
.
Voltages above 1200 V can damage the module.
Each module is tested with a test voltage of 5 kV
DC
for one min-
ute. A longer duration or a higher
voltage can damage the mod-
ule. In addition, each period of overvoltage reduces the service
life of the module.
Fig. 4-86 A127, measurement of voltage.
IMPORTANT
Voltages which exceed the admissible limits give incorrect mea-
surement data, because the input voltage is internally limited.
4.21.2 Current
Confusing the current and voltage inputs can damage the module
and/or the external load resistance.
Make sure that no high voltages are applied to the current inputs.
The voltage drop must not be beyond ±2.4 V.
You need an (external) shunt resistance for the current measu
re-
ment, refer also to Section 6.5, page 147.
Fig. 4-87 A127, current measurement using an external shunt
resistance.
U
2 (+)
4 (–)
I
R
ext.
7 (+)
9 (–)

Table of Contents

Related product manuals