GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A311 - AWS Unit
Page no. 1409 Chapter 9
JC-DR-A-311.fm
5. Reconnect the ground cable (screwdriver).
6. Reconnect all the cables to the AWS unit rear panel. Refer to ADS Rear Panel Connectors (Sun U20
Workstation) on page 1415 for a guide to cable connections.
6-4 Disassembly/Reassembly of AWS Unit on V4 Control Stations
6-4-1 Disassembly of AWS Unit
1. Locate the AWS unit (1) inside the Control Station.
2. Remove all the connected cables from the AWS unit rear panel.
Note:
Note locations of all disconnected cables.
3. Remove the ground cable (screwdriver).
4. Remove the two allen screws and associated bracket (2) securing the front of the AWS unit (4 mm
allen wrench).
1
2