GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
Page no. 627 Chapter 7
JC-ELE-A-021.fm
7 NETWORK PARAMETER CONFIGURATION
!
Notice:
When reloading system software from cold (i.e., including the operating system), first disconnect
the cable to the external network, and configure the system as "not networked". When the proce-
dure is complete, connect the cable and configure the system as "networked" before executing the
following procedure.
7-1 Enter Configuration Mode
1. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
2. Log in as root as follows:
a. At the ADS prompt enter:
su -
b. When requested to specify the root password, enter:
operator
3. Run the system configuration mode by entering:
sys-unconfig
4. At the warning message that appears, answer y (yes) to continue.
5. The AWS shuts down. Wait until AWS power has been switched off by the UPS (AWS shut down,
power light on monitor goes out). Then press the UPS ON switch on the Generator Cabinet to power
up the AWS.
6. The AWS starts up in configuration mode.
7. The system now displays a succession of screens in which you must enter the network parameters.
Specify the network parameters as outlined in the following sections. Where necessary, use the
arrow keys to move the cursor, and use the <Return> key to select an item. The screens order
depend on the type of Control Station you are configuring.
• For V1/V2 Control Stations, follow Network Configuration for V1/V2 Control Stations on page 628.
• For V3 Control Stations, follow Network Configuration for V3 Control Stations on page 634.
• For V4 Control Stations, follow Network Configuration for V4 Control Stations on page 643.

Table of Contents

Other manuals for GE Senographe DS