GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 857 Chapter 9
ERR-SUB-A-003.fm
102/013 HALT LIFT OVERSPEED The lift speed measured by the encoder is too high in
relation to the speed profile (and correlation is OK).
Refer to JC DIAG A1012 LIFT CONTROL FAILURE
on page 948 to fix the problem.
A1012
page 948
102/014 WARNING LIFT LOWSPEED The lift speed measured by the encoder is too low in
relation to the speed profile (and correlation is OK).
Check the lift operation in low speed and then refer
to JC DIAG A1013 LIFT CONTROL FAILURE on
page 948 to fix the problem.
A1013
page 948
102/015 HALT LIFT OVER CUR-
RENT
Occurs when an integrated value of the current is too
high, that is when the current is too high during too
much time.
Refer to JC DIAG A1011 LIFT CONTROL FAILURE
on page 947 to fix the problem.
A1011
page 947
102/016 DEBUG
ONLY
Movement already in
progress
Occurs when node is performing a movement, and
the Gantry CPU sends a command for another
movement when the first movement is not finished.
This is a CPU problem because the CPU did not wait
to send the next movement command.
Should not occur in the field. There could be an
incompatibility between the CPU software release
and the Node software release. Contact your OLC if
the problem persists.
102/017 HALT LIFT NODE FW
INCOMPATIBILITY
There is an Invalid Node Type. Occurs when the
hardware configuration (CPLD) of the node or the
DSP or the MCU firmware not match.
Change the Lift board then perform Lift Calibration.
D/R A201
page
1161
CAL A031
page
1687
102/018 HALT LIFT NODE PARITY
ERROR
There is a Parity error. Occurs when the parity bit
does not set correctly on the on-board bus between
the CPLD and the MCU.
Change the Lift board then perform Lift Calibration.
D/R A201
page
1161
CAL A031
page
1687
102/019 N/A STOP LINE WAS
ACTIVATED
BECAUSE OF
UNCONTROLLED
MOTION
Same as error 102/004.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card