5 - Quality Control(QC)
60 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
61 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
15 The test result is displayed (
Figure 67
).
Figure 67
: LQC Test Result
Pass – If the test passes, go to the next step.
Fail – If the test fails, review the package insert for LQC product. Run the test again with a new test
cartridge. If the test fails again, contact Technical Support.
16 Remove the test cartridge and dispose in a biohazardous waste container in accordance with
institutional guidelines.
BIOHAZARD: Test cartridges are for single-use only, do NOT reuse. All used test cartridges and
LQC products should be considered as potentially infectious. Strictly adhere to the institution’s
policies for the proper handling and disposal of biohazardous materials.
NOTE: The LQC test result will remain on the screen until a new test is performed or the user selects
(Figure 67).
NOTE: If Auto Send is enabled, the instrument will automatically send the test result to the network
and (Test Result Transmission Sent) will be displayed indicating the transmission was successful
(Figure 68).
Figure 68
: Test Result Transmission Sent