8 -  Centralized Conguration Manager(CCM) Software
88 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
89 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Setting Description
EQC Patient Tests
Maximum 
Patient Tests 
Allowed After 
EQC Expiration 
(0–240)
The number of additional patient tests that 
Operators
 are allowed to run after the 
EQC Interval expires.
Supervisors
 can allow 
Operators
 to run as many as 240 additional patient tests after the 
EQC Interval expires. When the maximum number of patient tests has been reached, if 
an 
Operator
 tries to run another patient test, QC lockout occurs.
To prevent 
Operators
 from running any additional patient tests, enter 
0
 (zero). As soon 
as the EQC Interval expires, QC lockout will occur.
When an additional patient test has been run, the value in the eld labeled 
“Number of 
Patient Tests Used After EQC Expiration”
 is incremented. 
Supervisors
 can clear this 
value by selecting Yes in the eld labeled 
“Clear Number of Patient Tests Used After 
EQC Expiration”
.
NOTE: It is recommended to run a successful EQC test to clear the value in the eld 
labeled “Number of Patient Tests Used After EQC Expiration”.
Number of 
Patient Tests 
Used After EQC 
Expiration
A counter that displays the number of additional patient tests that 
Operators
 have run 
after the EQC Interval expired. This value is read‑only.
After the EQC Interval expires, this counter is incremented each time an 
Operator
 runs 
an additional patient test.
For more information, see the description for the eld labeled 
“Maximum Patient Tests 
Allowed After EQC Expiration (0–240)”
.
Clear Number of 
Patient Tests 
Used After EQC 
Expiration
Choose whether to clear (reset) the counter that displays the number of patient tests 
that 
Operators
 have run after the EQC Interval expired:
•  No – Do not clear the counter.
•  Yes – Clear the counter.
When Yes is selected, the counter labeled 
“Number of Patient Tests Used After EQC 
Expiration”
 will be cleared (reset) when conguration data is sent to the instrument.
NOTE: It is recommended to run a successful EQC test to clear the value in the eld 
labeled “Number of Patient Tests Used After EQC Expiration”.