8 - Centralized Conguration Manager(CCM) Software
88 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
89 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Setting Description
EQC Patient Tests
Maximum
Patient Tests
Allowed After
EQC Expiration
(0–240)
The number of additional patient tests that
Operators
are allowed to run after the
EQC Interval expires.
Supervisors
can allow
Operators
to run as many as 240 additional patient tests after the
EQC Interval expires. When the maximum number of patient tests has been reached, if
an
Operator
tries to run another patient test, QC lockout occurs.
To prevent
Operators
from running any additional patient tests, enter
0
(zero). As soon
as the EQC Interval expires, QC lockout will occur.
When an additional patient test has been run, the value in the eld labeled
“Number of
Patient Tests Used After EQC Expiration”
is incremented.
Supervisors
can clear this
value by selecting Yes in the eld labeled
“Clear Number of Patient Tests Used After
EQC Expiration”
.
NOTE: It is recommended to run a successful EQC test to clear the value in the eld
labeled “Number of Patient Tests Used After EQC Expiration”.
Number of
Patient Tests
Used After EQC
Expiration
A counter that displays the number of additional patient tests that
Operators
have run
after the EQC Interval expired. This value is read‑only.
After the EQC Interval expires, this counter is incremented each time an
Operator
runs
an additional patient test.
For more information, see the description for the eld labeled
“Maximum Patient Tests
Allowed After EQC Expiration (0–240)”
.
Clear Number of
Patient Tests
Used After EQC
Expiration
Choose whether to clear (reset) the counter that displays the number of patient tests
that
Operators
have run after the EQC Interval expired:
• No – Do not clear the counter.
• Yes – Clear the counter.
When Yes is selected, the counter labeled
“Number of Patient Tests Used After EQC
Expiration”
will be cleared (reset) when conguration data is sent to the instrument.
NOTE: It is recommended to run a successful EQC test to clear the value in the eld
labeled “Number of Patient Tests Used After EQC Expiration”.