Step  1.  Turn the LCR-6000 Series ON. 
Step  2.  Set up the measurement conditions by filling in 
the fields on the MEAS DISPLAY page. 
  Move to the FUNC field using the cursor 
keys and choose Cs-D. 
  Move to the FREQ field using the cursor 
keys and input 1 kHz. 
  Move to the LEVEL field using the cursor 
keys and input 1 V   
Step 3.  Connect the test fixture to the LCR-6000 
Series. 
Step 4.  To compensate for the test fixture, a user 
correction is required.  
Don’t connect any DUT to the unit, as shown 
below: 
1.  Press the [Measure] key and the [OPEN 
SHORT] soft-key. 
2.  Move to the OPEN TEST or SPOT field 
using the cursor keys. 
3.  Don’t connect the clips to anything, as 
shown here: 
4.  Press the [MEAS OPEN] soft-key and 
then the [OK] soft-key. Wait until the 
“Correction finished” message is 
displayed. 
5.  Press the [ON] soft-key if the OPEN field 
is set to OFF. 