Column electric components
4990851_002_00 – 2082262 – 2022-01-10 67
21. Switch off the operating table using the switch at the running
gear.
22. Mount the sensor plate for the Trendelenburg end stops (2
screws).
23. Guide the column cover upwards and close (Chapter 6.6).
24. Mount the column head cover (Chapter 9.3).
25. Mount the pad plate onto the strut (Chapter 7.2).
26. Put the pad in place.
27. Connect the power supply to the operating table
(Chapter 6.2).
28. Update the firmware of the operating table.