12.4 Formulae
207
Total harmonic voltage distortion factor THDUF (%)
Single-phase 2-wire
1P2W
Single-phase 3-wire
1P3W
Three-phase 3-wire
3P3W2M
Three-phase 3-wire
3P3W3M
Three-phase 4-wire
3P4W
THDUF1
THDUF4
THDUF1
THDUF2
THDUF4
THDUF12
THDUF32
THDUF4
THDUF12
THDUF23
THDUF31
THDUF4
THDUF1
THDUF2
THDUF3
THDUF4
Calculate the Discrete Fourier Transform of harmonic RMS voltage at 2048 points (about once every 10
cycles at 50 Hz or every 12 cycles at 60 Hz).
Three-phase 3-wire indicates the result of harmonic calculation with line-to-line voltage.
In the equation above, K indicates the total of orders analyzed.
Select THDUF or THDUR to calculate the total harmonic voltage distortion factor.
c: measured channel, k: order for analysis
THDUFc
U
ck()
2
k2=
K
Uc1
-----------------------
100×=
Total harmonic current distortion factor THDIF (%)
Single-phase 2-wire
1P2W
Single-phase 3-wire
1P3W
Three-phase 3-wire
3P3W2M
Three-phase 3-wire
3P3W3M
Three-phase 4-wire
3P4W
THDIF1
THDIF4
THDIF1
THDIF2
THDIF4
THDIF1
THDIF2
THDIF4
THDIF1
THDIF2
THDIF3
THDIF4
THDIF1
THDIF2
THDIF3
THDIF4
Calculate the Discrete Fourier Transform of harmonic RMS current at 2048 points (about once every 10
cycles at 50 Hz or every 12 cycles at 60 Hz).
In the equation above, K indicates the total of orders analyzed.
Select either THDIF or THDIR to calculate the total harmonic current distortion factor.
c: measured channel, k: order for analysis
THDIFc
I
ck()
2
k2=
K
Ic1
---------------------
100×=
Total harmonic voltage distortion factor THDUR (%)
Single-phase 2-wire
1P2W
Single-phase 3-wire
1P3W
Three-phase 3-wire
3P3W2M
Three-phase 3-wire
3P3W3M
Three-phase 4-wire
3P4W
THDUR1
THDUR4
THDUR1
THDUR2
THDUR4
THDUR12
THDUR32
THDUR4
THDUR12
THDUR23
THDUR31
THDUR4
THDUR1
THDUR2
THDUR3
THDUR4
Calculate the Discrete Fourier Transform of harmonic RMS voltage at 2048 points (about once every 10
cycles at 50 Hz or every 12 cycles at 60 Hz).
Three-phase 3-wire indicates the result of harmonic calculation with line-to-line voltage.
In the equation above, K indicates the total of orders analyzed.
Select either THDUF or THDUR to calculate the total harmonic voltage distortion factor.
c: measured channel, k: order for analysis
THDURc
U
ck()
2
k2=
K
Uck()
2
k1=
K
-----------------------
100×=