EasyManuals Logo

HP 8360 L Series User Manual

HP 8360 L Series
429 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
4. While monitoring the internal detector, the RF modulator level
is varied until the detected power is equivalent to the reference
power measured in step 2.
5. Modulation is re-enabled if appropriate.
These steps are performed in approximately 200 ps and are repeated
any time power or frequency is changed.
See Also
Softkeys listed above, Fltness Menu, (MOD), [POWER LEVEL),
Set
Atten
“Externally Leveling the Swept CW Generator” in Chapter 1
“Working with Mixers/Reverse Power Effects” in Chapter 1
“Working with Spectrum Analyzers/Reverse Power Effects” in
Chapter 1
Operating and Programming Reference A-9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8360 L Series and is the answer not in the manual?

HP 8360 L Series Specifications

General IconGeneral
BrandHP
Model8360 L Series
CategoryInverter
LanguageEnglish

Related product manuals