Doubler
Amp
Mode Off
Programming Codes
SCPI:
POWer:AMPLifier:STATE ONlOFFlOll
POWer:AMPLifier:STATE?
Analyzer: NONE
See Also
Dblr Amp Menu
Doubler Amp Mode On
Function Group
POWER
Menu Map 5
Description
This softkey is applicable to instrument models with a doubler
installed. The doubler has an integral amplifier whose operation is
controlled by the instrument firmware. This softkey turns off the
automatic mode of operation and turns on the amplifier so that it is
always used. This is an unspecified mode of operation since it can
cause increased harmonics and degraded dynamic range at some
frequencies. This softkey has no effect on instruments without a
doubler.
An asterisk next to the key label indicates that this feature is active.
Programming Codes
SCPI:
POWer:AMPLifier:STATE
ONlOFFlOll
POWer:AMPLifier:STATE?
Analyzer: NONE
See Also
Dblr Amp Menu
D-8 Operating and Programming Reference