Save Lock
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
This softkey lets you disable the save function. It prohibits the
saving of the present instrument state into a save/recall memory
register. If this function is active, an error message is displayed. An
asterisk next to the key label indicates that this function is active.
SCPI: NONE
Analyzer: SHSV locks the registers, SHRC unlocks the registers.
(SAVE), Security Menu
“Saving and Recalling an Instrument State” in Chapter 1
“Programming Typical Measurements” in Chapter 1
Save User Preset
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
This softkey lets you store the present state of operation to be used
as the PRESET state. Set the swept CW generator to the desired
operating conditions. Select Save User Preset . The display shows:
--
>
User Defined Preset Saved
To activate this stored information, you must set the preset mode to
User.
SCPI: SYSTem:PRESet:SAVE
Analyzer: NONE
Preset Mode User
“Changing the Preset Parameters” in Chapter 1
S-2 Operating and Programming Reference