Types of Commands
Commands can be separated into two groups, common commands
and subsystem commands. Common commands are generally not
measurement related. They are used to manage macros, status
registers, synchronization, and data storage. Common commands
are easy to recognize because they all begin with an asterisk,
such as *IDN?,
*OPC,
and *RST. Common commands are defined
by IEEE 488.2. Subsystem commands include all measurement
functions and some general purpose functions. Subsystem commands
are distinguished by the colon used between keywords, as in
:
FREQuency
:CW?. Each command subsystem is a set of commands
that roughly corresponds to a functional block inside the instrument.
For example, the
POWer
subsystem contains commands for power
generation, while the STATUS subsystem contains commands for
accessing status registers.
l-l
SPCI
Common
Commands
“FIST
*IDN?
Subsystem
Commands
:MEAS:VOLT?
:FREQ
1KHz
Do75b
Figure l-24.
SCPI
Command Types
The remaining paragraphs in this subsection discuss subsystem
commands in more detail. Remember, some commands are
implemented in one instrument and not in another, depending on its
measurement function.
Getting Started Programming 1-67

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8360 L Series and is the answer not in the manual?

HP 8360 L Series Specifications

General IconGeneral
BrandHP
Model8360 L Series
CategoryInverter
LanguageEnglish

Related product manuals