Meter Adrs
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
In cases where the swept CW generator is capable of acting as a
controller to an HP 437B power meter, this softkey enables you to set
the programming address of the power meter. The address value can
be set from 0 to 30, with the factory default address set at 13. The
address value is stored in non-volatile memory.
SCPI: DIAGnostics:INSTrument:PMETer:ADDRess <num>
Analyzer: NONE
Adrss Menu
“Optimizing Swept CW Generator Performance” in Chapter 1
Chapter 3
(MOD)
Function Group
MENU SELECT
Menu Map
4
Description
This
hardkey
allows access to the pulse modulation function.
Pulse For use with Hewlett-Packard scalar analyzers, the
swept CW generator offers a scalar pulse modulation
mode that provides approximately 2 ps rise and fall
times. An internal oscillator provides the 27.778 kHz
square wave with no external connections necessary.
The slow waveform reduces the spectral width of the
output, improving measurements made on filters with
steep skirts.
See Also
Pulse On/Off Scalar
Operating and Programming Reference M-9