Freq Mult
Freq
Mult
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This softkey lets you set a frequency multiplier value and applies it to
all frequency parameters. Any integer value between and including
f36
is accepted. Changing the multiplier value changes the display,
it does not affect the output of the swept CW generator.
For example:
1. Set the start frequency to 4 GHz.
2. Set the stop frequency to 10 GHz.
3. Set the frequency multiplier to 5.
Note that the display indicates start=20 GHz, stop=50 GHz and
asterisks appear next to the frequency data.
4. Now set the stop frequency to 30 GHz. The swept CW generator
frequency is 6 GHz, or 30 GHz
+
5.
Frequency multiplier and offset are related as shown by the following
equation:
Entered value
or
Displayed Frequency = (Frequency Generated x Multiplier) + Offset value
The factory preset value is 1. An asterisk next to the key label
indicates that this feature is active.
SCPI:
FREQuency:MULTiplier
<num>lMAXimumlMinimum
FREQuency:MULTiplier:STATe
ONIOFF(l(0
<num> will be rounded to the nearest integer.
Analyzer: SHFA <n>
FREQUENCY
[MENU),
Freq Offset