Fault Info 2
Programming Codes
SCPI: NONE
Analyzer: NONE
See Also
Fault Menu
Fault Info 3
Function Group
SERVICE
Menu Map
6
Description
This softkey displays the latched status of the following fault
messages.
CALYO FAIL
Indicates that the YO adjusted at power-on or at
preset is unable to calibrate. Initiate a full self-test
to gather more information if this fault is indicated.
Indicates that the manual sweep DAC adjusted at
power-on or at preset is unable to calibrate. Initiate
a full self-test to gather more information if this fault
is indicated.
CALMAN
FAIL
TMR
CNFLCT
FAIL
SEARCH
Indicates a possible internal software error. Two
routines are trying to use the same timer.
Indicates that the ALC search leveling algorithm has
failed. This fault indication is possible only if the
search leveling mode is on.