Execution Errors are reported by the swept CW generator after
rounding and expression evaluation operations have taken place.
Errors that generate Execution Errors do not generate Command
Errors, Device-specific Errors, or Query Errors.
-240, Hardware error; Rear panel HP-IB switch
-224, Illegal parameter value
-222, Data out of range;Expected O-l
-222, Data out of range
-221, Settings conflict
-221, Settings
conflict;List
Arrays Invalid
-221, Settings conflict;Power And Level Mode
-221, Settings conflict;Power and attenuator
-221, Settings conflict;mm Module Mismatch
-220, Parameter
error;Value
not allowed
-213,
Init
ignored
-200, Execution
error;No
more room in EEROM
-200, Execution
error;Option
Not Installed
Error Messages From
-
199 to -100
These error messages indicate that a SCPI syntax error has been
detected by the swept CW generator’s parser. An error of this type
sets the Command Error Bit (bit 5) in the Event Status Register.
One of the following events has occurred:
1
A syntax error has been detected. Possible errors are: a data
element that violates the device listening formats or whose type is
unacceptable to the instrument.
n A semantic error has been detected indicating that an unrecognized
header was received.
n A Group Execute Trigger (GET)
was entered into the input buffer
inside a SCPI program message.
Events that generate Command Errors do not generate Execution
Errors, Device-specific Errors, or Query Errors.
-178, Expression data not allowed
-170, Expression
error;Bad
terminator
-161, Invalid block data;Bad terminator
-160, Block data error
-160, Block data
error;Bad
block type
-151, Invalid string data;Bad terminator
-144, Character data too
long;>12
chars
-141, Invalid character data;Bad char in token
-138, Suffix not allowed
-131, Invalid suffix;This one not allowed
-123, Exponent too 1arge;Decimal number
-123, Exponent too 1arge;Numeric overflow
-122, RESERVED
-121, Invalid character in number
-120, Numeric data
error;Bad
format
Error Messages
2a-7