Affinity
2.0
Additional Information Page 157
3) Mount the probe tube on the small microphone on the in-situ headset and place it on the ear of the
client using either the plastic headband or slide-on ear loop.
Plastic headband (click it on the plastic headband and hang it over the ears of the client).
Slide-on ear loop (slide on as shown below and hang it on the ear of the client).
The measurement tool on the headset allows for easy measurement of probe tube insertion depth
and the probe tube stabilizer can be used to hold the probe in place.
Measurement tool Probe tube stabilizer
Perform otoscopy and carefully insert the probe tube into the ear canal. Distance from ring to probe
tube tip should be approximately 31mm for males, 28mm for females, and 20-25 for children.
Attach the probe tube here