2.7.4.5 Measured RECD compared to a predicted RECD
You can at any time select a predicted RECD as described in section 0 and view it alongside a measured
RECD. Note that when both a measured and a predicted RECD are displayed the system will always
prioritize the measured RECD for the test box verification (simulated REM). If you wish to use the predicted
RECD the measured curve must be deleted.
2.7.4.6 Re-using a Coupler Curve
If you prefer to use a pre-recorded coupler curve this can be setup in the Settings menu for RECD in the
REM440 setup (see section 2.6.2.4).
When ticking Use prerecorded coupler curve the latest coupler curve will automatically appear on the graph
which means that the clinician only needs to perform the ear measure.
Note: This is only possible when using a Pure Tone stimulus.
Measured RECD
Predicted RECD