Additional Information Page 167
HA1 Mold refers using an RECD measured using the clients own ear mold followed by 2cc coupler
measures with an ITE adaptor.
HA1 Tip refers using an RECD measured using a foam tip or SPL probe followed by 2cc coupler
measures with an ITE adaptor.
HA2 Mold refers using an RECD measured using the clients own ear mold followed by 2cc coupler
measures with a BTE adaptor.
HA2 Tip refers using an RECD measured using a foam tip or SPL probe followed by 2cc coupler
measures with a BTE adaptor (the insert foam tip connects to the in-situ headset the same way as
described for HA2 Mold above).
Note: Remember to indicate the correct coupler type before performing the RECD based 2 cc coupler
measures.
(For more information about the DSL mi/o v5.0a please refer to www.dslio.com
) s