Additional Information Page 182
A predicted REUR may be selected if the REUR is a part of the individual test protocol (see section
2.6.2.1).
7) Carefully insert the hearing aid without moving the probe tube. Ensure that the instruments are
switched on.
8) If fitting an open hearing aid; perform Calibrate for open fit (see section 0).
9) Click on the REIG button if it is not already marked automatically.
A right click on REIG button allows you to enter the Settings menu or Change colour of the curve.
10) Check that the information about client, fitting prescription and hearing aid parameters to the right
are correct.
11) Ensure that the correct Ear is selected
12) Choose the preferred Stimulus in the corresponding dropdown list
If you have done Calibrate for open fit only the signals used for this calibration are available in the
stimulus selection dropdown (see section 0).
13) Select input level using the Input level slider
14) Click START and the REIG will be performed.
Ensure that the hearing aids are switched on and that the room is quiet whilst the curve is obtained.