Affinity
2.0
Additional Information Page 196
1) Open the REM440 module by selecting the REM tab.
2) If needed select a test protocol in the List of Protocols. If you do not select a protocol the system
will automatically select a standard protocol.
3) Explain the procedure to the client.
4) Perform otoscopic examination.
5) Mount the probe tube on the in-situ headset and place it on the client who should be seated ½-1
meters distance from the speaker.
Ensure that the probe tubes are calibrated before placing them in the ear canal (see section 0 about
tube calibration).
6) Carefully insert the hearing aid without moving the probe tube. Ensure that the instruments are
switched on.
7) If fitting an open hearing aid solution, perform Calibrate for open fit (see section 0).
8) Click on the Input/Output button if it is not already marked automatically.
A right click on the Input/Output button allows you to enter the Settings menu or Change colour of
the curve.
9) Ensure that the correct Ear is selected
10) Choose the preferred Stimulus in the corresponding dropdown list
If you have performed Calibrate for open fit only the signals used for this calibration are available in
the stimulus selection dropdown (see section 0).
11) Click START and the Input/Output measure will be performed.