EasyManua.ls Logo

Kamstrup MULTICAL 603 - Page 65

Kamstrup MULTICAL 603
155 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MULTICAL®603
KamstrupA/S∙Technicaldescription∙55122029_A1_GB_06.2017
65
t1error t2error
t3error V1Flowerror V2Flowerror
t1inlet Display‐‐‐
t2outlet Display‐‐‐
t(t1t2)
Display‐‐‐ Display‐‐‐
t3 Display‐‐‐
Flow,V1 
Power,V1 Display‐‐‐ Display‐‐‐
E1 Nocounting Nocounting
E2 Nocounting Nocounting
E3 Nocounting Nocounting
E4 Nocounting Nocounting
E5 Nocounting Nocounting
E6 Nocounting
E7 Nocounting Nocounting
E8 Nocounting
E9 Nocounting
E10 Nocounting
E11 Nocounting
V1 
V2 
A1 Nocounting Nocounting
A2 Nocounting Nocounting
Displayreadingincaseofcutoffofflow
Afterasystemcutoff,thecurrentflowreadinginMULTICAL®603willdropto0l/hduringonlyafewsecondswhena
flowsensorwithfastpulsessuchasULTRAFLOW®isused.
WhenMULTICAL®603isconnectedtoflow
sensorswithslowpulses,e.g.areedswitch,itwillfirstindicateadecreasing
currentflowseveralminutesafterthecutoff.MULTICAL®603automaticallysetstheflowreadingto0l/hafter60
minuteswithoutpulses.
Forflowsensorswithslowpulses,theflowreadingwillingeneralreactslower
andbelesssuitableforindicatinglow
flowsthanwhenusingflowsensorswithfastpulses.