Series 2280 Precision DC Power Supply Reference Manual Section 3: Functions and features
077085503 / March 2019 3-11
Using the front panel to set fill mode:
1. Press the MENU key.
2. Under Measure, select Data Buffers.
3. Select the button next to Mode. The Mode screen is displayed.
4. Select the buffer Mode. For example, press the button next to Always.
5. Press the HOME key to return to the home screen.
To set the reading buffer mode to overwrite old readings when the buffer is full, send the following
command:
:TRACe:FEED:CONTrol ALWays
To set the reading buffer mode to start but not overwrite old readings, send the following command:
:TRACe:FEED:CONTrol NEXT
To set the reading buffer mode so that no readings are saved in the buffer, send the following
command:
:TRACe:FEED:CONTrol NEVer
To set the reading buffer to hold 300 readings, send the following command:
:TRACe:POINts 300