Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-17
If a different function is selected, sending this command changes the measurement function to the
one specified by <function>. This function remains selected after the measurement is complete.
When this command is sent, the following commands execute in the following order:
ï‚§ :CONFigure[n]:<function>?
ï‚§ :READ[n]?
When :CONFigure is executed, the MEASure? parameters (<range> and <resolution>) are
executed and the instrument goes into the single-measurement mode.
When READ? is executed, the following commands execute in the order below:
ï‚§ :ABORt
ï‚§ :INITiate
ï‚§ :FETCh?
Example
:MEAS:CURR? 10,5
Measure current on the10 A range with 5.5 digit display
resolution.
Example output:
2.473375E-02A
Also see
:CONFigure[n]:<function> (on page 7-12)
:READ[n]? (on page 7-18)