Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-68 077085503 / March 2019
When you read this setting, you see the measurement range that the instrument is presently using. If
you change the range while the output is off, the instrument does not update the hardware settings,
but if you read the range setting, the return is the setting that will be used when the output is turned
on. If you set a range while the output is on, the new setting takes effect immediately. When you
assign a specific value to this command, the instrument is set on a fixed range that is large enough to
measure the assigned value. The instrument selects the best range for measuring a value of
<range>.
For example, for current measurements, if you expect a reading of approximately 50 mA, set
<range> to 0.05 (or 50e-3) to select the 100 mA range.
For concurrent measurements, range only applies to current measurement. For voltage
measurement, there is only one range.
Example
:SENS:CURR:RANG 0.01
Select the 10 mA range.
:SENS:CONC:RANG 1
Select the 1 A range.
:SENS:VOLT:RANG 32
Select the 32V range.
Also see
:SENSe[n]:<function>:RANGe:AUTO (on page 7-69)

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals