Models 2500 and 2502 User’s Manual Triggering 10-7
Bench defaults
The bench defaults are listed as follows. They are also denoted in Figure 10-1 by the “
symbol.
Arm-In Event = Immediate
•Trigger-In Source = Immediate
Arm Count = 1
•Trigger Count = 1
•Trigger Delay = 0.0 sec
Delay Action = 0.0005 sec
Source Trigger In Event = On
Delay Trigger In Event = Off
Measure Trigger In Event = Off
•Trigger Out Events = All Trigger Out Events are disabled (off)
Arm Out Events = Off
•Event Detection Bypasses = Never (both layers)
When the output is turned ON, the Model 2500 will run in a continuous loop around the
trigger model. After each Measure Action, operation will continue at the top of the trigger
model. The Model 2500 can be returned to idle by turning the output OFF.
Operation summary
The trigger model is designed to offer versatility for the various source and measure appli-
cations. Typically, it allows you to perform a specified number of measurements at various
source levels.
For example, assume you want to perform three measurements each at two different chan-
nel 1 voltage bias source levels (1V and 2V). To do this, set the arm count to two, the trig-
ger count to three, select SRC1, and use a 6-point Custom Sweep configured as follows:
P0000 = 1V
P0001 = 1V
P0002 = 1V
P0003 = 2V
P0004 = 2V
P0005 = 2V
When the sweep is started, operation falls into the Trigger Layer and performs three mea-
surements at the 1V source level. Operation then loops back into the Trigger Layer to per-
form three measurements at the 2V source level. The six readings are stored in the buffer.
Note that after the sweep is finished, the Model 2500 does not return to idle. Operation
continues at the top of the trigger model. Subsequent measurements are performed at the
2V level and are not stored in the buffer.
For details on the Custom Sweep, see Section 9.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals