EasyManua.ls Logo

Keithley 2500

Keithley 2500
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2500 and 2502 User’s Manual Connections 2-11
Analog output connections (Model 2502 only)
Analog output connector terminals
Figure 2-9 shows the terminal configuration of each triax ANALOG OUT connector.
Output terminals are as follows:
Center conductor: analog output HI
Inner shield: analog output LO (floating analog common)
Outer shield: chassis ground
WARNING Analog output low can float up to ±100V. Use care to avoid a shock
hazard when using the analog outputs.
Figure 2-9
Analog output connector terminals
Analog Output HI
Analog Output LO
(Analog Common)
Chassis Ground
ANALOG OUT
Triax Connector
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Related product manuals