Models 2500 and 2502 User’s Manual SCPI Command Reference 17-41
MODE <name>
:DISPlay:MODE <name> Select CALC block for display
Parameters <name> = CALCulate3 Display CALC3 (Ch. 1 REL)
CALCulate4 Display CALC4 (Ch. 2 REL)
CALCulate5 Display CALC5 (RATIO)
CALCulate6 Display CALC6 (DELTA)
DUAL Dual-channel display mode
Query :MODE? Query selected display CALC block
Description This command is used to select which CALC block to display. Dis-
played CALC blocks include CALC3 (channel 1 REL), CALC4 (chan-
nel 2 REL), CALC5 (RATIO), and CALC6 (DELTA). The DUAL
parameter selects the dual-channel display mode.
ATTRibutes?
:DISPlay[:WINDow[1]]:ATTRibutes? Query attributes; top display
:DISPlay:WINDow2:ATTRibutes? Query attributes; bottom display
Description This query command is used to determine which characters on the dis-
play are blinking and which are not. The response message provides that
status of each character position for the specified display. The primary
display consists of 20 characters and the secondary display consists of
32 characters.
1 = Character is blinking
0 = Character is not blinking
For example, assume the following menu is displayed with the
SAVESETUP option blinking:
MAIN MENU
SAVESETUP COMMUNICATION CAL >
The response message for :disp:attr? (top display) will result in 20
zeroes as follows:
00000000000000000000
The response message for :disp:wind2:attr? (bottom display) will dis-
play ones at the character positions for SAVESETUP as follows:
11111111100000000000000000000000
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com