17-88 SCPI Command Reference Models 2500 and 2502 User’s Manual
Configure and control buffer
FREE?
:TRACe:FREE? Read status of memory
Description This command is used to read the status of storage memory. After send-
ing this command and addressing the Model 2500 to talk, two values
separated by commas are sent to the computer. The first value indicates
how many bytes of memory are available, and the second value indi-
cates how many bytes are reserved to store readings.
POINts <n>
:TRACe:POINts <n> Specify buffer size
Parameters <n> = 1 to 3000 Specify buffer size
MINimum 1
MAXimum 3000
DEFault 100
Query :POINts? Query buffer size
:POINts? MINimum Query smallest allowable buffer size
:POINts? MAXimum Query largest allowable buffer size
:POINts? DEFault Query *RST default buffer size
Description This command is used to specify the size of the buffer.
ACTual?
:TRACe:POINts:ACTual? Query number of stored readings
Description This query command is used to determine how many stored readings are
in the buffer. After sending this command and addressing the unit to
talk, the number of readings stored in the buffer will be sent to the com-
puter.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com