Models 2500 and 2502 User’s Manual SCPI Command Reference 17-5
Command Description
Default
parameter SCPI
:CALCulate7
:DATA?
:LATest?
:FEED <name>
:FEED?
:LIMit[1]
:COMPliance
:FAIL <name>
:FAIL?
:SOURce3 <NRf> | <NDN>
:SOURce3?
:STATe <b>
:STATe?
:FAIL?
:LIMit2
:COMPliance
:FAIL <name>
:FAIL?
:SOURce3 <NRf> | <NDN>
:SOURce3?
:STATe <b>
:STATe?
:FAIL?
Path to CALC7 (limit tests):
Return all CALC7 results.
Return most recent CALC7 result.
Specify Limit 3 to 6 input path (CALCulate3,
CALCulate4, CALCulate5, or CALCulate6).
Query input path.
Path to channel 1 hardware (compliance limits).
Specify fail mode (IN or OUT) of compliance.
Specify output fail pattern (0 to 7 | 15).
1
Query fail bit pattern.
2
Specify IN or OUT pattern (0 to 7| 15).
1
Query IN or OUT pattern.
2
Enable or disable Limit 1 test.
Query state of Limit 1 test.
Returns result of Limit 1 test: 0 (pass) or 1 (fail).
Path to channel 2 hardware (compliance) limit.
Specify fail mode (IN or OUT) of compliance.
Specify output fail pattern (0 to 7 | 15).
1
Query fail bit pattern.
2
Specify IN or OUT pattern (0 to 7| 15).
1
Query IN or OUT pattern.
2
Enable or disable Limit 2 test.
Query state of Limit 2 test.
Returns result of Limit 2 test: 0 (pass) or 1 (fail).
IN
15 or 7
OFF
IN
15 or 7
OFF
✓
✓
✓
✓
✓
✓
1
Default based on present digital output size (:SOURce3:BSIZe <n>). 3-bit default is 7. 4-bit default is 15. 16-bit default is 65535.
(16-bit available only with 2499-DIGIO option.)
2
The format (ASCII, hexadecimal, octal, or binary) for the returned value is set by FORMat:SOURce3 <name>.
Table 17-1 (continued)
CALCulate command summary
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176