Loading...
Models 2500 and 2502 User’s Manual Basic Operation 3-9
Figure 3-3
Ground connect enabled
Figure 3-4
Ground connect disabled
Floating
Common
To Ammeter
Bias Source
SOURCE
OUTPUT
Triax
INPUT
Note: One channel shown.
Chassis
Ground
HI
LO
Ground Connect
Relay Closed
Chassis
Ground
Output
On
DUT
Other channel is identical.
Floating
Common
To Ammeter
Bias Source
SOURCE
OUTPUT
Triax
INPUT
Note: One channel shown.
Chassis
Ground
HI
LO
Ground
Connect
Relay
Open
Chassis
Ground
Output
On
DUT
Other channel is identical.