17-56 SCPI Command Reference Models 2500 and 2502 User’s Manual
Set measurement speed
NPLCycles <n>
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n> Set channel 1 speed (PLC)
:SENSe2:CURRent[:DC]:NPLCycles <n> Set channel 2 speed (PLC)
Parameters <n> = 0.01 to 10 Power-line cycles per integration
DEFault 1
MINimum 0.01
MAXimum 10
Query :NPLCycles? Query programmed PLC value
:NPLCycles? DEFault Query *RST default PLC
:NPLCycles? MINimum Query minimum PLC
:NPLCycles? MAXimum Query maximum PLC
Description This command is used to set the integration period (speed) for measure-
ments. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC
of 1, the integration period would be 1/60 (for 60Hz line power) which
is 16.67 msec.
Note that this is a global command. Thus, if you set the speed for chan-
nel 1 measurements to 10 PLC, then channel 2 will also set to 10 PLC.