EasyManuals Logo

Keithley 2500 User Manual

Keithley 2500
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
Models 2500 and 2502 User’s Manual Measurement Concepts 5-7
Regardless of the load, current will never exceed the fixed compliance of 20mA.
Figure 5-5
Loading effects
A. Normal bias source operation
B. Bias source in compliance
V
S
= I
M
R
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
20mA
10V
= (20mA) (400)
= 8V
8V
400 DUT Load Line (R)
Output
Current (I)
Output Voltage (V)
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
Output Voltage (V)
Output
Current (I)
10mA
20mA
10V
1k DUT Load Line (R)
I = V / R
= 10V/1k
= 10mA
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals