17-80 SCPI Command Reference Models 2500 and 2502 User’s Manual
ENABle <list>
:STATus:QUEue:ENABle <list> Enable messages for error queue
Parameters <list> = (numlist)
where numlist is a specified list of messages that you wish to enable for
the Error.
Query :ENABle? Query list of enabled messages
Description On power-up, all error messages are enabled and will go into the error
queue as they occur. Status messages are not enabled and will not go
into the queue. This command is used to specify which messages you
want enabled. Messages not specified will be disabled and prevented
from entering the queue.
DISable <list>
:STATus:QUEue:DISable <list> Disable messages for error queue
Parameters <list> = (numlist)
where numlist is a specified list of messages that you wish to disable for
the error queue.
Query :DISable? Query list of disabled messages
Description On power-up, all error messages are enabled and will go into the error
queue as they occur. Status messages are not enabled and will not go
into the queue. This command is used to specify which messages you
want disabled. Disabled messages are prevented from going into the
error queue.