1-10 Getting Started Model 2500 and 2502 User’s Manual
Power-up
Line voltage selection
The Model 2500 operates from a line voltage in the range of 100V/120V/220V/240V,
±10% at a frequency of 50 or 60Hz. Before plugging in the unit make sure the line voltage
setting seen through the small window in the power module (Figure 1-2) is correct for the
line voltage in your area.
CAUTION Operating the instrument on an incorrect line voltage may cause dam-
age, possibly voiding the warranty.
If the voltage setting is not correct, change it as outlined below.
WARNING Disconnect the line cord and all cables and test leads from the instru-
ment before changing the line voltage setting.
1. Pry the small tab that secures the fuse holder to the power module, then remove the
fuse holder.
2. Pull out and rotate the fuse carrier to the correct line voltage setting, then install it in
the fuse holder. Make sure the line voltage shown in the window is correct. Also make
sure the fuse rating for the expected line voltage is correct. (See “Fuse replacement.)
3. Reinstall the fuse holder in the power module making sure it is seated fully.
Line power connection
Perform the following steps to connect the Model 2500 to line power and turn it on:
1. Before plugging in the power cord make sure the front panel power switch is in the off
(0) position.
2. Connect the female end of the supplied power cord to the AC receptacle on the rear
panel.
WARNING The power cord supplied with the Model 2500 contains a separate
ground for use with grounded outlets. When proper connections are
made, instrument chassis is connected to power line ground through
the ground wire in the power cord. Failure to use a grounded outlet
may result in personal injury or death due to electric shock.
3. Turn on the instrument by pressing the front panel power switch to the on (1) position.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals