Models 2500 and 2502 User’s Manual SCPI Command Reference 17-61
Select sourcing mode
MODE <name>
:SOURce[1]:VOLTage:MODE <name> Select sourcing mode for source 1
:SOURce2:VOLTage:MODE <name> Select sourcing mode for source 2
Parameters <name> = FIXed Select fixed sourcing mode
LIST Select list sourcing mode
SWEep Select sweep sourcing mode
Query :MODE? Query DC sourcing mode
Description These commands are used to select the sourcing mode for the specified
source. The three modes are explained as follows:
FIXed — In this sourcing mode, the specified source will output a fixed
level. Use the :RANGe and :AMPLitude commands to specify the fixed
source level. See “Select range,” page 17-61, and “Set amplitude for
fixed source,” page 17-63.
LIST — In this mode, the source will output levels that are specified in
a list. See “Configure list,” page 17-71, for commands to define and con-
trol the execution of the list.
SWEep — In this mode, the source will perform a voltage sweep. See
“Configure sweeps,” page 17-66, for commands to define the sweep.
NOTE The sourcing mode will default to FIXed whenever the Model 2500 goes to the
local state.
Select range
RANGe <n>
:SOURce[1]:VOLTage:RANGe <n> Select range for channel 1 source
:SOURce2:VOLTage:RANGe <n> Select range for channel 2 source
Parameters <n> = -100 to 100 Specify source level (volts)
DEFault 10V range
MINimum 10V range
MAXimum 100V range
UP Select next higher range
DOWN Select next lower range
Query :RANGe? Query range for specified source
:RANGe? DEFault Query *RST default source range
:RANGe? MINimum Query lowest source range
:RANGe? MAXimum Query highest source range