Models 2500 and 2502 User’s Manual SCPI Command Reference 17-93
source occurs immediately. Subsequent arm events will then occur every
30 seconds. The interval for the timer is set using the :TIMer command.
With MANual selected, the event occurs when the TRIG key is pressed.
With BUS selected, the event occurs when a GET or *TRG command is
sent over the bus.
With NSTest selected, the event occurs when the SOT (start-of-test) low
pulse is received from a component handler via the digital I/O port. This
is used for limit testing.
With PSTest selected, the event occurs when SOT (start-of-test) high
pulse is received from a component handler via the digital I/O port. This
is used for limit testing.
With BSTest selected, the event occurs when SOT (start-of-test) high or
low pulse is received from a component handler via the digital I/O port.
This is used for limit testing.
TIMer <n>
:ARM[:SEQuence[1]][:LAYer[1]]:TIMer <n> Set interval for arm layer timer
Parameters <n> = 0.001 to 9999.999 Specify timer interval in seconds
10000.00 to 99999.99 Specify timer interval in seconds
Query :TIMer? Query programmed timer interval
Description These commands are used to set the interval for the timer. Note that the
timer is in effect only if the timer is the selected control source.
DIRection <name>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:DIRection <name> Control arm bypass
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection <name> Control trigger bypass
Parameters <name> = SOURce Enable control source bypass
ACCeptor Disable control source bypass
Query :DIRection? Query state of bypass
Description This command is used to enable (SOURce) or disable (ACCeptor) con-
trol source bypass. When enabled, operation will loop around the con-
trol source on the first pass in the layer. After that, repeat passes in the
layer are held up and will wait for the programmed control source event.