Models 2500 and 2502 User’s Manual SCPI Command Reference 17-35
PASS:SOURce3 <NRf> | NDN
:CALCulate7:LIMitx:PASS:SOURce3 <NRf> | <NDN> Set sorting mode pass pattern (x = 3-6)
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce3? Query programmed source value
Description This command is used to define the 3-bit or 4-bit output pattern for the
digital I/O port when a test (Limit 3-6) for the sorting mode passes. Note
that the output value can be specified in binary, octal, decimal, or hexa-
decimal format. Use the table provided in “SOURce3 <NRf> |
<NDN>, “Description,” page 17-34, to determine the parameter value
for the desired decimal digital output pattern.
NOTE 16-bit I/O is available with the 2499-DIGIO option. The maximum 16-bit output
value is 65535.
STATe <b>
:CALCulate7:LIMit[1]:STATe <b> Control Limit 1 (channel 1) test
:CALCulate7:LIMit2:STATe <b> Control Limit 2 (channel 2) test
:CALCulate7:LIMitx:STATe <b> Control Limit x test (x = 3-6)
Parameters <b> = 1 or ON Enable specified limit test
0 or OFF Disable specified limit test
Query :STATe? Query state of specified limit test
Description These commands are used to enable or disable Limit 1 through Limit 6,
tests. Limit 1 and Limit 2 are the compliance limit tests for channel 1
and channel 2 respectively, while Limit 3 through Limit 6 are the soft-
ware limit tests. Any limit test not enabled is simply not performed.
When a limit test is enabled, the Digital I/O port comes under control of
limit tests. That is, the result of the testing process updates the output
pattern on the I/O port.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals