17-32 SCPI Command Reference Models 2500 and 2502 User’s Manual
Read limits data
DATA?
:CALCulate7:DATA? Read limits data
Description This command is used to acquire all the readings used for the CALC7
limit tests. At least one of the limit tests must be enabled to acquire limit
test readings. See “Configure and control limit tests;” “STATe <b>,”
page 17-35.
LATest?
:CALCulate7:DATA:LATest? Read latest CALC7 data
Description This command works exactly like CALC7:DATA?, except it returns
only the latest limit result.
Configure and control limit tests
COMPliance:FAIL <name>
:CALCulate7:LIMit[1]:COMPliance:FAIL <name> Set channel 1 compliance fail mode
:CALCulate7:LIMit2:COMPliance:FAIL <name> Set channel 2 compliance fail mode
Parameters <name> = IN Fail Limit 1/2 test when unit goes into
compliance
OUT Fail Limit 1/2 test when unit comes out
of compliance
Query :FAIL? Query when Limit 1/2 test failure occurs
Description These commands are used to specify the condition that will cause the
Limit 1 and Limit 2 tests to fail. With IN specified, the test will fail
when the Model 2500 goes into compliance (fixed 20mA current limit
for each voltage source). With OUT specified, the test will fail when the
Model 2500 comes out of compliance.
[:DATA] <n>
:CALCulate7:LIMitx:LOWer[:DATA] <n> Specify lower Limit x (x = 3-6)
:CALCulate7:LIMitx:UPPer[:DATA] <n> Specify upper Limit x (x = 3-6)
Parameters <n> = -9.999999e20 to Specify limit value
9.999999e20
DEFault Set specified lower limit to -1
Set specified upper limit to 1
MINimum Set specified limit to -9.999999e20
MAXimum Set specified limit to +9.999999e20