Models 2500 and 2502 User’s Manual SCPI Command Reference 17-81
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
Table 17-8.
Default conditions
PRESet
:SYSTem:PRESet Return to :SYSTem:PRESet defaults
Description This command returns the instrument to states optimized for front panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables
(Table 17-1 through Table 17-10).
POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet defaults
SAV0 Power-up to setup stored at memory location 0
SAV1 Power-up to setup stored at memory location 1
SAV2 Power-up to setup stored at memory location 2
SAV3 Power-up to setup stored at memory location 3
SAV4 Power-up to setup stored at memory location 4
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the SCPI tables
(Table 17-1 through Table 17-10).
With the SAV0-4 parameters specified, the instrument powers-on to the
setup that is saved in the specified location using the *SAV command.