Models 2500 and 2502 User’s Manual SCPI Command Reference 17-59
Median filter commands
MEDian:RANK <NRf>
[:SENSe[1]]:MEDian:RANK <n> Set median filter rank
:SENSe2:MEDian:RANK <n> Set median filter rank
Parameters <n> = 0 to 5 Specify rank value for median filter
DEFault 1
MINimum 0
MAXimum 5
Query RANK? Query median filter rank
Description The median filter is used to pass the middle-most reading from a group
of readings that are arranged according to size. This command is used to
specify the rank, which determines the number of reading samples for
the filter process.
The number of reading samples are determined as follows:
Sample readings = 2n + 1
Where: n is the selected rank (1 to 5).
MEDian[:STATe] <b>
[:SENSe[1]]:MEDian[:STATe] <b> Enable/disable channel 1 median filter
:SENSe2:MEDian[:STATe] <b> Enable/disable channel 2 median filter
Parameters <b> = 0 or OFF Disable repeat filter
1 or ON Enable repeat filter
Query [:STATe]? Query state of repeat filter
Description This command is used to enable or disable the median filter. When
enabled, current readings are filtered according to the specified rank.
When disabled, the median filter stage is bypassed.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals