Models 2500 and 2502 User’s Manual Sweep Operation 9-15
Custom sweep programming example
As an example of custom sweep operation, assume a five-point sweep with the following
parameters:
Source and measure channel: channel 1
Voltage sweep mode: list (custom sweep)
Sweep voltage points: 7V, 1V, 3V, 8V, 2V
Source delay: 100ms
Table 9-5 summarizes the basic remote command sequence for performing the custom
sweep described above.
Table 9-5
Custom sweep programming example
Command Description
*RST
:FORM:ELEM CURR1
:SOUR1:VOLT:MODE LIST
:SOUR1:LIST:VOLT 7,1,3,8,2
:TRIG:COUN 5
:SOUR1:DEL 0.1
:OUTP1 ON
:READ?
:OUTP1 OFF
Restore GPIB default conditions.
Select channel 1 measurement data.
Channel 1 list volts sweep mode.
7V, 1V, 3V, 8V, 2V Ch. 1 sweep points.
Trigger count = # sweep points.
100ms Ch. 1 source delay.
Turn on Ch. 1 source output.
Trigger sweep, request data.
Output 1 off.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals