Models 2500 and 2502 User’s Manual SCPI Command Reference 17-57
Configure and control filters
NOTE Detailed information on the average and median filters are provided in
Section 6, “Filters.”
All filter commands are global and commands affect both channels
simultaneously.
Average filter commands
COUNt <n>
[:SENSe[1]]:AVERage:COUNt <n> Set average filter count
:SENSe2:AVERage:COUNt <n> Set average filter count
Parameters <n> = 1 to 100 Specify average filter count
DEFault 10
MINimum 1
MAXimum 100
Query :COUNt? Query filter count
:COUNt? DEFault Query the *RST default filter count
:COUNt? MINimum Query the lowest allowable filter count
:COUNt? MAXimum Query the largest allowable filter count
Description These commands are used to specify the average filter count for channel
1 and channel 2. In general, the filter count is the number of readings
that are acquired and stored in the filter buffer for the averaging calcula-
tion. Each aquired group of readings yields a single filtered reading. The
larger the filter count, the more filtering that is performed.
[:STATe] <b>
[:SENSe[1]]:AVERage:[:STATe] <b> Enable/disable average filter
:SENSe2:AVERage:[:STATe] <b> Enable/disable average filter
Parameters <b> = 0 or OFF Disable average filter
1 or ON Enable average filter
Query [:STATe]? Query state of average filter
Description These commands are used to enable or disable the average filter. When
enabled, current readings are filtered according to how the average filter
is configured (see TCONtrol <name>). When disabled, the average filter
stage is bypassed.