17-24 SCPI Command Reference Models 2500 and 2502 User’s Manual
Set MX + B parameters
MBFactor <n>
:CALCulate[1]:KMATh:MBFactor <n> Select channel 1 B (offset) value
:CALCulate2:KMATh:MBFactor <n> Select channel 2 B (offset) value
Parameters <n> = -9.99999e20 to Specify B (offset) value for MX + B
+9.99999e20
MINimum -9.99999e20
MAXimum +9.99999e20
DEFault 0
Query :MBFactor? Query B (offset) value for MX + B
Description These commands program the B (offset) value for the MX + B math
function (see FORMat <name> above). Use CALC1 for channel 1 or
CALC2 for channel 2.
MMFactor <n>
:CALCulate[1]:KMATh:MMFactor <n> Select channel 1 M (slope) value
:CALCulate2:KMATh:MMFactor <n> Select channel 2 M (slope) value
Parameters <n> = -9.99999e20 to Specify M (slope) value for MX + B
+9.99999e20
MINimum -9.99999e20
MAXimum +9.99999e20
DEFault 1
Query :MMFactor? Query M (slope) value for MX + B
Description These commands program the M (slope) value for the MX + B math
function (see FORMat <name> above). Use CALC1 for channel 1 or
CALC2 for channel 2.
MUNits <name>
:CALCulate[1]:KMATh:MUNits <name> Specify units for channel 1 MX + B
:CALCulate2:KMATh:MUNits <name> Specify units for channel 2 MX + B
Parameters <name> = Up to three ASCII characters enclosed in single or double
quotes
Query :MUNits? Query units for MX + B
Description These commands are used to specify the units suffix name for the MX +
B math function for channel 1 (CALC1) and channel 2 (CALC2). Use
up to three ASCII characters for the units suffix name consisting of
upper- or lower-case characters (no numbers, spaces, dashes, etc.).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176